us_artefacts
Ultrasound image artefacts
Introduction:
US systems operate on a basis of certain false assumptions relating to the interaction of the sound beam with soft tissue, including:
a constant speed of sound in the body of 1540m/s
all echoes detected by the transducer originate from the central axis of the beam
the ultrasound beam travels in straight lines
the time taken for an echo from a given interface to return to the transducer is directly related to its distance from the transducer
the rate of attenuation of the beam is constant with depth & throughout the field of view
as a result, artefacts may be produced such as:
echoes in the display which do not appear in the correct position
absence of echoes from parts of the image that would normally generate echoes
structures which are anatomically separate being displayed as joined
brightness of echoes which may not correspond to that expected
single structure being displayed as two separate structures
thin membranes appearing thicker than they really are
Acoustic shadowing:
if the proportion of the beam energy attenuated at a given interface is high, little sound will be transmitted deeper than the interface, thus echoes will not be received from tissue deep to the interface resulting in acoustic shadowing
examples:
Acoustic enhancement:
an attenuation phenomenon resulting in an area of increased brightness relative to echoes from adjacent tissue
this is due to sound passing through a medium of lower sound attenuation than the adjacent tissue resulting in the sound reaching the tissues deep to it having greater amplitude than the sound reaching the adjacent tissues as the same depth
examples:
the margins of the area of enhancement are usually sharply defined & may be parallel or divergent depending on the transducer used
Edge shadowing:
sound is attenuated deep to the lateral edges of rounded structures such as aorta
the sound has further to travel within that medium resulting absorption as well as reflection & refraction effects in producing the narrow band of distal edge shadow
Reverberation artefact:
Beam width effects:
Velocity artefacts:
Mirror image:
Operator produced artefacts:
excessive gain:
poor TGC settings:
Problem solving:
turn down the gain
alter angulation of probe and watch effect on the artefact
scan in two planes
move patient
have patient use different breathing manoeuvres
change transducer frequency
Next:
us_artefacts.txt · Last modified: 2008/11/19 23:45 (external edit)