ventpsv
Table of Contents
Pressure Support Ventilation (PSV)
see also:
introduction
- PSV is a flow-limited mode of ventilation that delivers inspiratory pressure until the inspiratory flow decreases to a predetermined percentage of its peak value. This is usually 25%.
- During PSV, each breath is augmented by inspiratory pressure.
- The clinician sets the pressure support level (inspiratory pressure level), applied PEEP, and FiO2
- There is no set respiratory rate, thus, the patient triggers each inspiration, which continues until the inspiratory flow decreases to a system-specific minimal level.
- Exhalation follows.
- The tidal volume is determined by the pressure support level, effort, and mechanics.
- Minute volume is variable and dependent upon tidal volume and patient initiated respiratory rate, and thus does not guarantee an adequate minute volume.
- increasing the level of pressure support decreases the work of breathing provided that inspiratory flow is sufficient to meet patient demand.
- PSV seems particularly well suited for weaning from mechanical ventilation because it tends to be a comfortable mode, giving the patient greater control over the inspiratory flow rate and respiratory rate.
- PSV is frequently combined with SIMV. The ventilator delivers the set respiratory rate using SIMV, but patient-initiated breaths beyond the set respiratory rate are delivered using PSV. The purpose of adding PSV for patient-initiated breaths is to overcome the resistance of the endotracheal tube and ventilator circuit.
- in general, PSV is poorly suited to provide full or nearly full ventilatory support, it is a particularly poor choice for patients who also have increased airway resistance (eg, chronic obstructive pulmonary disease (COPD) or asthma exacerbation).
pressure control ventilation (PCV)
- machine breathes are pressure limited and either time triggered or patient triggered
- ventilator delivers flow to achieve a preset level of pressure early in inspiration
- decelerating flow pattern
ventilator settings
- frequency
- upper and lower level of pressure
- inspiratory time
- thus flow adjusts automatically
- tidal volume is determined by:
- pressure gradient - the difference between the set PIP and PEEP, and also patient effort
- inspiratory time
- patient and machine characteristics including lung and chest wall compliance, airway resistance, and circuit characteristics
advantages
- improved oxygenation via:
- higher mean airway pressure
- better gas flow distribution
- spontaneous breathing at upper level (if allowed)
- able to meet high inspiratory flow demands
- better patient and ventilator synchrony in some modes
- less sedation needs
- active exhalation valve
- can breathe at any stage during the cycle
Pressure Support Ventilation (PSV)
- applies to spontaneous breathing only
- maintains and supports the patient's inspiratory effort
- used as a mode on its own or often combined with SIMV
- flow cycled
- once synchronisation is detected, a constant flow is delivered at the preset pressure
- expiration is triggered by a reduction in the inspiratory flow to a given level
advantages
- improved patient and ventilator synchrony
- increased patient comfort
- minimal or no sedation may be required
- minimal cardiovascular effects
- patient retains control over frequency, inspiratory flow and time
disadvantages
- minute volume is not guaranteed
- large changes in patients breathing may occur without changing the actual minute volume
- periods of hypoventilation may occur
- increased work of breathing if inappropriate rise time or trigger settings
- patient may fail to cycle to expiration if leak
- failure to detect apnoea if use of an inline nebuliser
uses
- used to overcome ETT and circuit resistance encountered during spontaneous breathing
- used to augment tidal volume in spontaneous breathes
- reduces work of breathing
- to assess if extubation is likely to be successful
- weaning from intubation
ventpsv.txt · Last modified: 2014/09/29 03:38 by 127.0.0.1